A geometric calibration method for the digital chest tomosynthesis with dual-axis scanning geometry

一种用于双轴扫描几何结构数字胸部断层合成的几何校准方法

阅读:2

Abstract

The aim of this study was to develop a geometric calibration method capable of eliminating the reconstruction artifacts of geometric misalignments in a tomosynthesis prototype with dual-axis scanning geometry. The potential scenarios of geometric misalignments were demonstrated, and their effects on reconstructed images were also evaluated. This method was a phantom-based approach with iterative optimization, and the calibration phantom was designed for specific tomosynthesis scanning geometry. The phantom was used to calculate a set of geometric parameters from each projection, and these parameters were then used to evaluate the geometric misalignments of the dual-axis scanning-geometry prototype. The simulated results revealed that the extracted geometric parameters were similar to the input values and that the artifacts of reconstructed images were minimized due to geometric calibration. Additionally, experimental chest phantom imaging results also indicated that the artifacts of the reconstructed images were suppressed and that object structures were preserved through calibration. And the quantitative analysis result also indicated that the MTF can be further improved with the geometric calibration. All the simulated and experimental results demonstrated that this method is effective for tomosynthesis with dual-axis scanning geometry. Furthermore, this geometric calibration method can also be applied to other tomography imaging systems to reduce geometric misalignments and be used for different geometric calibration phantom configurations.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。