Comparative endurance testing of the Biomet Matthews Nail and the dynamic compression screw, in simulated condylar and supracondylar femoral fractures

在模拟股骨髁部和髁上骨折中,对 Biomet Matthews 髓内钉和动态加压螺钉进行耐久性对比测试

阅读:1

Abstract

BACKGROUND: The dynamic compression screw is a plate and screws implant used to treat fractures of the distal femur. The Biomet Matthews Nail is a new retrograde intramedullary nail designed as an alternative surgical option to treat these fractures. The objective of this study was to assess the comparative endurance of both devices. METHOD: The dynamic compression screw (DCS) and Biomet Matthews Nail (BMN) were implanted into composite femurs, which were subsequently cyclically loaded using a materials testing machine. Simulated fractures were applied to each femur prior to the application of load. Either a Y type fracture or a transverse osteotomy was prepared on each composite femur using a jig to enable consistent positioning of cuts. RESULTS: The Biomet Matthews Nail demonstrated a greater endurance limit load over the dynamic compression screw in both fracture configurations. CONCLUSION: The distal locking screws pass through the Biomet Matthews Nail in a unique "cruciate" orientation. This allows for greater purchase in the bone of the femoral condyle and potentially improves the stability of the fracture fixation. As these fractures are usually in weak osteoporotic bone, the Biomet Matthews Nail represents a favourable surgical option in these patients.