The Correlation between Carotid Artery Corrected Flow Time and Velocity Time Integral during Central Blood Volume Loss and Resuscitation

中心血容量丢失和复苏期间颈动脉校正血流时间与速度时间积分的相关性

阅读:1

Abstract

BACKGROUND: Doppler ultrasound of the common carotid artery is used to infer central hemodynamics. For example, change in the common carotid artery corrected flow time (ccFT) and velocity time integral (VTI) are proposed surrogates of changing stroke volume. However, conflicting data exist which may be due to inadequate beat sample size and measurement variability - both intrinsic to handheld systems. In this brief communication, we determined the correlation between changing ccFT and carotid VTI during progressively severe central blood volume loss and resuscitation. METHODS: Measurements were obtained through a novel, wireless, wearable Doppler ultrasound system. Sixteen participants (ages of 18-40 years with no previous medical history) were studied across 25 lower body-negative pressure protocols. Relationships were assessed using repeated-measures correlation regression models. RESULTS: In total, 33,110 cardiac cycles comprise this analysis; repeated-measures correlation showed a strong, linear relationship between ccFT and VTI. The strength of the ccFT-VTI relationship was dependent on the number of consecutively averaged cardiac cycles (R(1 cycle) = 0.70, R(2 cycles) = 0.74, and R(10 cycles) = 0.81). CONCLUSIONS: These results positively support future clinical investigations employing common carotid artery Doppler as a surrogate for central hemodynamics.