Validity of Wearable Gait Analysis System for Measuring Lower-Limb Kinematics during Timed Up and Go Test

可穿戴步态分析系统在计时起立行走测试中测量下肢运动学的有效性

阅读:1

Abstract

Few studies have reported on the validity of a sensor-based lower-limb kinematics evaluation during the timed up and go (TUG) test. This study aimed to determine the validity of a wearable gait sensor system for measuring lower-limb kinematics during the TUG test. Ten young healthy participants were enrolled, and lower-limb kinematics during the TUG test were assessed using a wearable gait sensor system and a standard optical motion analysis system. The angular velocities of the hip, knee, and ankle joints in sit-to-stand and turn-to-sit phases were significantly correlated between the two motion analysis systems (R = 0.612-0.937). The peak angles and ranges of motion of hip, knee, and ankle joints in the walking-out and walking-in phases were also correlated in both systems (R = 0.528-0.924). These results indicate that the wearable gait sensor system is useful for evaluating lower-limb kinematics not only during gait, but also during the TUG test.