Development of a multi-wavelength diffuse optical tomography system for early diagnosis of rheumatoid arthritis: simulation, phantoms and healthy human studies

用于类风湿性关节炎早期诊断的多波长漫射光学断层扫描系统的开发:模拟、体模和健康人体研究

阅读:1

Abstract

A multi-wavelength diffuse optical tomography (DOT) system has been developed to directly extract physiological information, such as total haemoglobin concentration, from tissue in human hand joints. Novel methods for 3D surface imaging and spectrally constrained image reconstruction are introduced and their potential application to imaging of rheumatoid arthritis is discussed. Results are presented from simulation studies as well as experiments using phantoms and data from imaging of three healthy volunteers. The image features are recovered partially for phantom data using transmission measurements only. Images that reveal joint regions and surrounding features within the hand are shown to co-register with co-acquired ultrasound images which are shown to be related to total haemoglobin concentration.