An In-Vitro Optical Sensor Designed to Estimate Glycated Hemoglobin Levels

用于估计糖化血红蛋白水平的体外光学传感器

阅读:15
作者:Sanghamitra Mandal, M O Manasreh

Abstract

The purpose of this research was to design an optical sensor for evaluating glycated hemoglobin (HbA1c) percentages in hemoglobin. The A1c sensors available in the market use invasive methods, while our device offers the possibility of non-invasive monitoring of HbA1c levels in diabetic patients. A prototype is assembled using two light emitting diodes with peak emission wavelengths of 535 nm and 593 nm, a photodiode, and a microcontroller. The proposed sensor measures the transmitted intensity in the form of an output voltage. We devise an approach to estimate the percentage of HbA1c in hemoglobin for a given solution. This estimation is based on the relative change in absorbance due to change in path length and molar absorption coefficients of hemoglobin and HbA1c, at the two wavelengths. We calculate the molar absorption coefficient of HbA1c at 535 nm and 593 nm wavelengths using the sensor, which is performed by a multiple variable regression analysis algorithm fed through the microcontroller. Specifically, the sensor output voltage with respect to the sample concentration is fitted to an exponentially decaying equation model. We used a commercial chemical assay called Control FD Glycohemoglobin A1c with known percentage HbA1c levels to verify our device measurements.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。