Safe Tissue Manipulation in Retinal Microsurgery via Motorized Instruments with Force Sensing

利用力感应电动器械在视网膜显微手术中安全操作组织

阅读:1

Abstract

Retinal microsurgery involves careful manipulation of delicate tissues by applying very small amount of forces most of which lie below the tactile sensory threshold of the surgeons. Membrane peeling is a common task in this domain, where application of excessive peeling forces can easily lead to serious complications, hence needs to be avoided. To quantify tool-tissue interaction forces during retinal microsurgery, various force-sensing tools were developed based on fiber Bragg grating sensors, yet the most beneficial way of using the acquired force information is currently unknown. In this study, using a motorized force-sensing micro-forceps tool, we develop an assistive method that enhances safety during membrane peeling by automatically opening the forceps and releasing the tissue based on the detected peeling forces. Through peeling experiments using bandages, we demonstrate that our method can effectively maintain the peeling force at a safe level even in case of non-homogeneous adhesion properties of the membrane.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。