Common-path phase-sensitive optical coherence tomography provides enhanced phase stability and detection sensitivity for dynamic elastography

共光路相位敏感光学相干断层扫描技术为动态弹性成像提供了更高的相位稳定性和检测灵敏度。

阅读:1

Abstract

Phase-sensitive optical coherence elastography (PhS-OCE) is an emerging optical technique to quantify soft-tissue biomechanical properties. We implemented a common-path OCT design to enhance displacement sensitivity and optical phase stability for dynamic elastography imaging. The background phase stability was greater in common-path PhS-OCE (0.24 ± 0.07nm) than conventional PhS-OCE (1.60 ± 0.11μm). The coefficient of variation for surface displacement measurements using conventional PhS-OCE averaged 11% versus 2% for common-path PhS-OCE. Young's modulus estimates showed good precision (95% CIs) for tissue phantoms: 24.96 ± 2.18kPa (1% agar), 49.69 ± 4.87kPa (1.5% agar), and 116.08 ± 12.14kPa (2% agar), respectively. Common-path PhS-OCE effectively reduced the amplitude of background dynamic optical phase instability to a sub-nanometer level, which provided a larger dynamic detection range and higher detection sensitivity for surface displacement measurements than conventional PhS-OCE.