Rapid identification of Ebola virus and related filoviruses in fluid specimens using indirect immunoelectron microscopy

利用间接免疫电镜技术快速鉴定液体标本中的埃博拉病毒及相关丝状病毒

阅读:1

Abstract

Recent filoviral outbreaks in animal primates have raised public awareness of the potential for filoviruses to become a public health concern; methods that efficiently identify these viruses are therefore of high priority. An indirect immunoelectron microscopy method, which uses homologous guinea pig polyclonal antiserum, successfully identified Ebola-related (Reston) virus particles in serum and tissue culture fluid specimens with infectivity titres of 300 plaque forming units (pfu) per ml or more. The sensitivity of this procedure is sufficient to show virus in most acute phase sera, and is equal to that of the antigen capture enzyme linked immunosorbent assay (ELISA). The immunoelectron microscopy fluid technique can differentiate among antigenically distinct filoviruses in less than three hours. It should be valuable in the rapid diagnosis of potential filoviral infections.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。