3D Dosimetry Based on LiMgPO(4) OSL Silicone Foils: Facilitating the Verification of Eye-Ball Cancer Proton Radiotherapy

基于LiMgPO(4) OSL硅箔的三维剂量测定:促进眼球癌质子放射治疗的验证

阅读:1

Abstract

A direct verification of the three-dimensional (3D) proton clinical treatment plan prepared for tumor in the eyeball, using the Eclipse Ocular Proton Planning system (by Varian Medical Systems), has been presented. To achieve this, a prototype of the innovative two-dimensional (2D) circular silicone foils, made of a polymer with the embedded optically stimulated luminescence (OSL) material in powder form (LiMgPO(4)), and a self-developed optical imaging system, consisting of an illuminating light source and a high-sensitive CCD camera has been applied. A specially designed lifelike eyeball phantom has been used, constructed from 40 flat sheet LMP-based silicone foils stacked and placed together behind a spherical phantom made by polystyrene, all to reflect the curvature of the real eyeball. Two-dimensional OSL signals were captured and further analyzed from each single silicone foil after irradiation using a dedicated patient collimator and a 58.8 MeV modulated proton beam. The reconstructed 3D proton depth dose distribution matches very well with the clinical treatment plan, allowing for the consideration of the new OSL system for further 3D dosimetry applications within the proton radiotherapy area.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。