A new plan quality objective function for determining optimal collimator combinations in prostate cancer treatment with stereotactic body radiation therapy using CyberKnife

本文提出了一种新的计划质量目标函数,用于确定使用CyberKnife立体定向放射治疗前列腺癌的最佳准直器组合。

阅读:1

Abstract

Stereotactic body radiation therapy with CyberKnife for prostate cancer has long treatment times compared with conventional radiotherapy. This arises the need for designing treatment plans with short execution times. We propose an objective function for plan quality evaluation, which was used to determine an optimal combination between small and large collimators based on short treatment times and clinically acceptable dose distributions. Data from 11 prostate cancer patients were used. For each patient, 20 plans were created based on all combinations between one small (⌀ 10-25 mm) and one large (⌀ 35-60 mm) Iris collimator size. The objective function was assigned to each combination as a penalty, such that plans with low penalties were considered superior. This function considered the achievement of dosimetric planning goals, tumor control probability, normal tissue complication probability, relative seriality parameter, and treatment time. Two methods were used to determine the optimal combination. First, we constructed heat maps representing the mean penalty values and standard deviations of the plans created for each collimator combination. The combination giving a plan with the smallest mean penalty and standard deviation was considered optimal. Second, we created two groups of superior plans: group A plans were selected by histogram analysis and group B plans were selected by choosing the plan with the lowest penalty from each patient. In both groups, the most used small and large collimators were assumed to represent the optimal combination. The optimal combinations obtained from the heat maps included the 25 mm as a small collimator, giving small/large collimator sizes of 25/35, 25/40, 25/50, and 25/60 mm. The superior-group analysis indicated that 25/50 mm was the optimal combination. The optimal Iris combination for prostate cancer treatment using CyberKnife was determined to be a collimator size between 25 mm (small) and 50 mm (large).

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。