A nanostructured piezoelectric immunosensor for detection of human cardiac troponin T

用于检测人源心肌肌钙蛋白T的纳米结构压电免疫传感器

阅读:1

Abstract

A piezoelectric immunosensor based on gold nanoparticles (AuNPs) co-immobilized on a dithiol-modified surface is proposed for detection of human cardiac troponin T (TnT). Anti-human troponin T (anti-TnT) antibodies were covalently immobilized on the nanostructured electrode surface by thiol-aldehyde linkages. In a homogeneous bulk solution, TnT was captured by anti-TnT immobilized on the QCM electrode. Cyclic voltammetry studies were used to characterize the AuNPs layer on the electrode surface and the anti-TnT immobilization steps. The QCM-flow immunosensor exhibited good reliability, measuring concentrations of TnT from 0.003 to 0.5 ng mL(-1) in human serum with high linearity (r = 0.989; p < 0.01). The immunosensor exhibited a 7% coefficient of variation and 0.0015 ng mL(-1) limit of detection, indicating a high reproducibility and sensitivity. The proposed QCM nanostructured immunosensor is easy to use and has promising potential in the diagnosis of acute myocardial infarction due to its speed and high sensitivity.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。