Commissioning Validation of a Brachytherapy Treatment Planning System with EGSnrc Monte Carlo Code and EBT3 GAFChromic Film

使用EGSnrc蒙特卡罗代码和EBT3 GAFChromic胶片对近距离放射治疗计划系统进行调试验证

阅读:1

Abstract

AIMS: Most brachytherapy treatment planning system (TPS) commissioning requires data input based on the American Association of Physicists in Medicine Task Group-43 formalism. The commissioning accuracy is very important for dose calculation. The aim of this study is the implementation of a brachytherapy TPS into a clinical environment and check the TPS calculated dose accuracy. SUBJECTS AND METHODS: After introducing data of the different catheters (CIS Bio International, Saclay, France), composed of several Cesium-137 Eckert and Ziegler BEBIG CSM-11 radioactive sources; for XiO (CMS, St. Louis) brachytherapy TPS, the TPS dose calculation accuracy was investigated by comparing between the TPS calculated dose distribution (DD) for all the catheters with (1) the measuring DD using EBT3 GAFChromic film and (2) calculating DD by egs_brachy (Electron Gamma Shower, National Research Council of Canada) Monte Carlo simulation. The phantom used for this study consists of six PTW slabs 30 cm × 30 cm × 1 cm of polymethyl methacrylate with a Delouche MEDpro applicator on the top. The TPS DD was calculated on the computed tomography scan of this phantom. STATISTICAL ANALYSIS USED: PTW VeriSoft version 6.0.1.7 (PTW-Freiburg, Germany) software was used for analyzing scanned films and to perform the comparison based on the gamma index distribution. RESULTS: For each catheter, the gamma index distribution showed agreement >95% of all pixels in both verification methods, with gamma ≤1. CONCLUSIONS: We confirm the commissioning accuracy and that the TPS can be used for clinical purposes.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。