A Novel Vascular Anastomotic Coupling Device for End-to-End Anastomosis of Arteries and Veins

用于动脉和静脉端端吻合的新型血管吻合耦合装置

阅读:2

Abstract

OBJECTIVE: Hand-sutured (HS) techniques remain the gold standard for most microvascular anastomoses in microsurgery. HS techniques can result in endothelial lacerations and back wall suturing, leading to complications such as thrombosis and free tissue loss. A novel force-interference-fit vascular coupling device (FIF-VCD) system can potentially reduce the need for HS and improve end-to-end anastomosis. This study aims to describe the development and testing of a novel FIF-VCD system for 1.5 to 4.0 mm outside diameter arteries and veins. METHODS: Benchtop anastomoses were performed using porcine cadaver arteries and veins. Decoupling force and anastomotic leakage were tested under simulated worst-case intravital physiological conditions. The 1.5 mm FIF-VCD system was used to perform cadaver rat abdominal aorta anastomoses. RESULTS: Benchtop testing showed that the vessels coupled with the FIF-VCD system could withstand simulated worst-case intravital physiological conditions with a 95% confidence interval for the average decoupling force safety factor of 8.2 ± 1.0 (5.2 ± 1.0 N) and a 95% confidence interval for the average leakage rate safety factor of 26 ± 3.6 (8.4 ± 0.14 and 95 ± 1.4 μL/s at 150 and 360 mmHg, respectively) when compared to HS anastomotic leakage rates (310 ± 14 and 2,100 ± 72 μL/s at 150 and 360 mmHg, respectively). The FIF-VCD system was successful in performing cadaver rat abdominal aorta anastomoses. CONCLUSION: The FIF-VCD system can potentially replace HS in microsurgery, allowing the safe and effective connection of arteries and veins. Further studies are needed to confirm the clinical viability and effectiveness of the FIF-VCD system.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。