Interaction between shock coils increased the incidence of inappropriate therapies and lead failure in implantable cardioverter defibrillator

植入式心脏复律除颤器中,电击线圈之间的相互作用增加了不适当治疗和导线故障的发生率。

阅读:1

Abstract

AIMS: Shock coil interaction in patients with multiple implantable cardioverter defibrillator (ICD) leads is occasionally observed. We aimed to evaluate the incidence of shock coil interaction and its clinical relevance. METHODS AND RESULTS: All ICD patients (646 patients) who came to follow up control in our ICD ambulance between January 1, 2011, and December 31, 2011 in the department of cardiology in Bad Berka hospital were retrospectively evaluated in this study. All baseline demographic, clinical, and procedural characteristics and postoperative chest x ray in postero-anterior and lateral view as well as clinical and ICD follow up data were evaluated. Among 646 patients 42 had multiple ICD leads (6.5%) of whom 36 patients (5.5% of total cohort patients and 85.7% of patients with multiple ICD leads) had shock coil interaction and presented the study group (Group I). The control group (Group II) consisted of 610 patients without coil-coil interaction including patients with single shock lead (604 patients) or patients with multiple leads but without interaction between shock coils (6 patients). Inappropriate anti-tachycardia therapies and RV lead revisions were more frequent in patients with interaction between shock coils (Group I vs Group II: 27.7% and 5.7%; p = 0.049 and 30.6% vs 6.4; p = 0.0001, respectively). CONCLUSIONS: Interaction between shock coils may be one of possible causes of lead failure and resulted in inappropriate therapies and subsequent lead revision.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。