Refractoriness to subcutaneous implantable cardioverter defibrillator after frequent therapies for ventricular fibrillation storms in a Brugada syndrome case

布鲁加达综合征病例中,在多次接受室颤风暴治疗后,皮下植入式心脏复律除颤器无效

阅读:1

Abstract

BACKGROUND: The subcutaneous implantable cardioverter defibrillator (S-ICD) is an alternative to the transvenous implantable cardioverter defibrillator for the prevention of sudden cardiac death. Here, we report a rare case of refractoriness to an S-ICD after frequent therapies for ventricular fibrillation (VF) storms. CASE PRESENTATION: A 24-year-old man underwent a bout of syncope with vomiting and incontinence at home. He was brought to the emergency room and was witnessed to spontaneously go into VF successfully converted by external defibrillation. Previously, he was diagnosed with a type I Brugada electrocardiogram pattern by a pilsicainide administration test in another hospital. Although he had a family history of sudden cardiac death in 3 relatives, including his brother, he was followed closely without any therapies because he had never had an episode of syncope. He was implanted with an S-ICD without any trouble. Seven months later, frequent S-ICD shocks for VF storms occurred. His VF was controlled by using intravenous amiodarone, which was converted to an oral preparation. However, his VF recurred after another 2 months. The analysis of his S-ICD data revealed that 4 consecutive shock deliveries could not terminate his VF and the final shock delivered could fortunately terminate it because of a high defibrillation threshold test (DFT) due to an increasing shock impedance (64 to 90 Ω). First, we performed an epicardial Brugada syndrome ablation and subsequently replaced and repositioned the S-ICD lead from a left to a right parasternal site. After the re-implantation of the S-ICD, the DFT test improved to within normal range. According to the pathological analysis, infiltration of inflammatory cells and extensive fibrosis were confirmed in the subcutaneous tissue around the shock lead and S-ICD body. CONCLUSION: Frequent S-ICD shocks for VF storms might cause various pathological changes around the device and lead to a high DFT.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。