The effect of intraoral suction on oxygen-enriched surgical environments: a mechanism for reducing the risk of surgical fires

口腔内吸力对富氧手术环境的影响:降低手术火灾风险的一种机制

阅读:1

Abstract

In this study, a mechanical model was applied in order to replicate potential surgical fire conditions in an oxygen-enriched environment with and without high-volume suction typical for dental surgical applications. During 41 trials, 3 combustion events were measured: an audible pop, a visible flash of light, and full ignition. In at least 11 of 21 trials without suction, all 3 conditions were observed, sometimes with an extent of fire that required early termination of the experimental trial. By contrast, in 18 of 20 with-suction trials, ignition did not occur at all, and in the 2 cases where ignition did occur, the fire was qualitatively a much smaller, candle-like flame. Statistically comparing these 3 combustion events in the no-suction versus with-suction trials, ignition (P = .0005), audible pop (P = .0211), and flash (P = .0092) were all significantly more likely in the no-suction condition. These results suggest a possible significant and new element to be added to existing surgical fire safety protocols toward making surgical fires the "never-events" they should be.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。