General‐Purpose Ultrasound Neuromodulation System for Chronic, Closed‐Loop Preclinical Studies in Freely Behaving Rodents (Adv. Sci. 34/2022)

用于自由活动啮齿动物慢性闭环临床前研究的通用超声神经调控系统(Adv. Sci. 34/2022)

阅读:1

Abstract

Neurogenic bladder is a dysfunction in the lower urinary tract due to damage to the nervous system. One of the treatments that has shown important results is transcutaneous neuromodulation. The neuromodulation equipment available on the market does not allow remote activation or offer many resources for adjusting the parameters of the generated stimulus, as most devices operate with pre-established parameters in closed packages. For this reason, customizing therapy for each individual can be difficult. Therefore, the objective was to develop and validate a neuromodulation device capable of being remotely programmed and properly monitored. Materials and methods for validating devices were used according to the Brazilian Regulatory Standard (NBR), which deals with general requirements for the basic safety and essential performance of electromedical devices. After verifying the reliability of the device, which was capable of generating a biphasic and symmetrical wave, measured by an oscilloscope, considered safe by the technical requirements, it was tested in a real application. Users reported feeling a comfortable stimulus, similar to other previously used devices, and considered the device easy to use. In this way, it was possible to demonstrate the reliability and validity of the developed device.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。