Comparison of automated and manual nucleic acid extraction methods for detection of enterovirus RNA

比较自动化和手动核酸提取方法在肠道病毒RNA检测中的应用

阅读:2

Abstract

Automated nucleic acid extraction is an attractive alternative to labor-intensive manual methods. We compared two automated methods, the BioRobot M48 instrument (Qiagen, Inc.) and MagNA Pure (Roche Applied Sciences) methods, to two manual methods, the QIAamp Viral RNA Mini kit (Qiagen) and TRIzol (Invitrogen), for the extraction of enterovirus RNA. Analytical sensitivity was assessed by dilution analysis of poliovirus type 2 Sabin in cerebrospinal fluid. The sensitivity of PCR was equivalent after RNA extraction with QIAamp, BioRobot M48, and MagNA Pure. All 18 replicates of 100 PFU/ml were detected after extraction by the four methods. Fewer replicates of each successive dilution were detected after extraction by each method. At 10(-1) PFU/ml, 17 of 18 replicates were positive by QIAamp, 15 of 18 replicates were positive by BioRobot M48, and 12 of 18 replicates were positive by MagNA Pure; at 10(-2) PFU/ml, 4 of 17 replicates were positive by QIAamp, 2 of 18 replicates were positive by BioRobot M48, and 0 of 18 replicates were positive by MagNA Pure. At 10(-3) PFU/ml, no replicates were detected. Evaluation of TRIzol was discontinued after nine replicates due to a trend of lower sensitivity (at 10(-3) PFU/ml eight of nine replicates were positive at 100 PFU/ml, four of nine replicates were positive at 10(-1) PFU/ml, and zero of nine replicates were positive at 10(-2) PFU/ml). Concordant results were obtained in 24 of 28 clinical specimens after extraction by all methods. No evidence of contamination was observed after extraction by automated instruments. The data indicate that the sensitivity of enterovirus PCR is largely similar after extraction by QIAamp, BioRobot M48, and MagNA Pure; a trend of decreased sensitivity was observed after TRIzol extraction. However, the results of enterovirus PCR were largely concordant in patient samples, indicating that the four extraction methods are suitable for detection of enteroviruses in clinical specimens.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。