Long-Term Evaluation of Bone Healing Monitoring Using an Instrumented Plate with Measurement Sensors (Smart Implant) over 10 Years

利用带有测量传感器的仪器化钢板(智能植入物)进行超过10年的骨愈合监测长期评估

阅读:1

Abstract

A total of 66 smart implants were included. As a measure of bony stability, the relative elastic compliance of the osteosynthesis was determined from the gradient between the applied external load and the measured implant load over the entire healing process. The healing process of non-unions of the femur with a smart implant was tracked by telemetric measurements over a timespan of up to 10 years. The measurements of the longest healing process show a very slow but constant decrease in force transmission over the implant, radiological findings over 10 years show corresponding consolidation until bony healing. The use of a telemetrically instrumented bone plate, a so-called smart implant, to monitor the healing process is a successful procedure to support the clinician in his decision to take further surgical measures or to wait until healing occurs.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。