Direct frequency domain fluorescence lifetime imaging using simultaneous ultraviolet and visible excitation

同时使用紫外和可见光激发的直接频域荧光寿命成像

阅读:9
作者:Michael J Serafino, Javier A Jo

Abstract

Due to the complexity, limited practicality, and cost of conventional fluorescence lifetime imaging/microscopy (FLIM) instrumentation, FLIM adoption has been mostly limited to academic settings. We present a novel point scanning frequency-domain (FD) FLIM instrumentation design capable of simultaneous multi-wavelength excitation, simultaneous multispectral detection, and sub-nanosecond to nanosecond fluorescence lifetime estimation. Fluorescence excitation is implemented using intensity-modulated CW diode lasers that are available in a selection of wavelengths spanning the UV-VI-NIR range (375-1064 nm). Digital laser intensity modulation was adopted to enable simultaneous frequency interrogation at the fundamental frequency and corresponding harmonics. Time-resolved fluorescence detection is implemented using low-cost, fixed-gain, narrow bandwidth (100 MHz) avalanche photodiodes, thus, enabling cost-effective fluorescence lifetime measurements at multiple emission spectral bands simultaneously. Synchronized laser modulation and fluorescence signal digitization (250 MHz) is implemented using a common field-programmable gate array (FPGA). This synchronization reduces temporal jitter, which simplifies instrumentation, system calibration, and data processing. The FPGA also allows for the implementation of the real-time processing of the fluorescence emission phase and modulation at up to 13 modulation frequencies (processing rate matching the sampling rate of 250 MHz). Rigorous validation experiments have demonstrated the capabilities of this novel FD-FLIM implementation to accurately measure fluorescence lifetimes in the range of 0.5-12 ns. In vivo endogenous, dual-excitation (375nm/445nm), multispectral (four bands) FD-FLIM imaging of human skin and oral mucosa at 12.5 kHz pixel rate and room-light conditions was also successfully demonstrated. This versatile, simple, compact, and cost-effective FD-FLIM implementation will facilitate the clinical translation of FLIM imaging and microscopy.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。