Radiation Dose Reduction in Cancer Imaging with New-Model CT Scanners: A Quality of Care Evaluation

利用新型CT扫描仪降低癌症影像检查中的辐射剂量:一项医疗质量评估

阅读:1

Abstract

BACKGROUND/OBJECTIVES: The primary aim of this study was to evaluate whether the replacement of roughly one-decade-old computed tomography (CT) scanners with new-model CT scanners were associated with an additional reduction in the radiation dose delivered to oncological patients, in a radiological setting where the optimization of protocols had already reached very low radiation doses. An exploratory secondary objective was to evaluate the potential differences in the objective image quality between the CT scans obtained before and after the installation of the new-generation CT scanners. METHODS: Chest and abdominal CT examinations conducted for oncologic purposes were retrospectively selected from two time periods-prior to scanner replacement (2022) and following an upgrade (2024)-after five CT systems in our radiology department were replaced. We extracted and compared the CT dose index (CTDI) and dose length product (DLP) for each CT phase. For the objective image quality evaluation, we calculated the signal-to-noise ratio (SNR) and the contrast-to-noise ratio (CNR) at the center of the liver and the aorta. An appropriate statistical analysis was performed and a p-value < 0.05 was considered significant. RESULTS: We included 14,601 CT acquisitions, of which 9013 (61.7%) were performed before and 5588 (38.3%) after the replacement of the CT scanners. There were significantly lower values for the CTDI and DLP with the new CT scanners compared to the old ones. The CTDI with the new CT scanners was significantly lower in all phases (p-value = 0.002 for unenhanced phase, and p < 0.001 for arterial, portal venous, and delayed phases). The DLP using the new CT scanners was significantly lower in the arterial, portal venous, and delayed phases (p < 0.001), and it was not significantly different in the unenhanced phase (p = 0.36). There was no significant difference in the SNR at the liver level (p = 0.72) or at the aorta level (p = 0.51). There was no significant difference in the CNR at the liver level (p = 0.24), whereas the CNR was higher with the new CT scanners at the aorta level (p = 0.03). CONCLUSIONS: The transition to new-model CT scanners resulted in a significant reduction in the radiation dose delivered by chest and abdomen CT scans, without compromising the objective image quality.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。