An Efficient and Multi-Focal Focused Ultrasound Technique for Harmonic Motion Imaging

一种用于谐波运动成像的高效多焦点聚焦超声技术

阅读:1

Abstract

Harmonic motion imaging (HMI) is an ultrasound-based elasticity imaging technique that utilizes oscillatory acoustic radiation force to estimate the mechanical properties of tissues, as well as monitor high-intensity focused ultrasound (HIFU) treatment. Conventionally, in HMI, a focused ultrasound (FUS) transducer generates oscillatory tissue displacements, and an imaging transducer acquires channel data for displacement estimation, with each transducer being driven with a separate system. The fixed position of the FUS focal spot requires mechanical translation of the transducers, which can be a time-consuming and challenging procedure. In this study, we developed and characterized a new HMI system with a multi-element FUS transducer with the capability of electronic focal steering of ±5 mm and ±2 mm from the geometric focus in the axial and lateral directions, respectively. A pulse sequence was developed to drive both the FUS and imaging transducers using a single ultrasound data acquisition (DAQ) system. The setup was validated on a tissue-mimicking phantom with embedded inclusions. Integrating beam steering with the mechanical translation of the transducers resulted in a consistent high contrast-to-noise ratio (CNR) for the inclusions with Young's moduli of 22 and 44 kPa within a 5-kPa background while the data acquisition speed is increased by 4.5-5.2-fold compared to the case when only mechanical movements were applied. The feasibility of simultaneous generation of multiple foci and tracking the induced displacements is demonstrated in phantoms for applications where imaging or treatment of a larger region is needed. Moreover, preliminary feasibility is shown in a human subject with a breast tumor, where the mean HMI displacement within the tumor was about 4 times lower than that within perilesional tissues. The proposed HMI system facilitates data acquisition in terms of flexibility and speed and can be potentially used in the clinic for breast cancer imaging and treatment.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。