Compensation for Electrode Detachment in Electrical Impedance Tomography with Wearable Textile Electrodes

利用可穿戴纺织电极进行电阻抗断层扫描时电极脱落的补偿

阅读:2

Abstract

Electrical impedance tomography (EIT) is a radiation-free and noninvasive medical image reconstruction technique in which a current is injected and the reflected voltage is received through electrodes. EIT electrodes require good connection with the skin for data acquisition and image reconstruction. However, detached electrodes are a common occurrence and cause measurement errors in EIT clinical applications. To address these issues, in this study, we proposed a method for detecting faulty electrodes using the differential voltage value of the detached electrode in an EIT system. Additionally, we proposed the voltage-replace and voltage-shift methods to compensate for invalid data from the faulty electrodes. In this study, we present the simulation, experimental, and in vivo chest results of our proposed methods to verify and evaluate the feasibility of this approach.