Dumbbell-shaped thrombectomy device for cerebral venous sinus thrombus removal with controllable axial and longitudinal maneuverability

用于脑静脉窦血栓切除的哑铃形血栓切除装置,具有可控的轴向和纵向机动性

阅读:1

Abstract

Cerebral venous sinus thrombosis (CVST) is frequently observed in younger adults and features in large thrombus volume. Due to the triangular-like cross-sectional shape and large diameter of the superior sagittal sinus, all the commercially available artery stent retrievers are not suitable for venous vessels. In this study, a dumbbell-like stent was designed and fabricated by 3D braided technology using NiTi wires; it was manually rotatable and stretchable with controlled length/diameter ratios (2.6-14.0) and reciprocating maneuverability. Computational modeling and an in vitro study were conducted to evaluate the mechanical properties of this device and its ability to trap and remove thrombi from occluded venous vessels was verified by using a swine model. A single-center retrospective clinical study of 10 patients using the Venus-TD to treat patients with CVST was also conducted. Pre/postoperative thrombus volume in 10 patients was quantitatively analysed (12 855.3 ± 6417.1 vs. 2373.1 ± 2759.0 mm³, P < 0.001) with a high recanalization rate, yielding favorable clinical outcomes. This study offers a novel treatment option for patients with extensive CVST.