Noninvasive Flow Monitoring in Simple Flow Phantom Using Resistive Strain Sensors

使用电阻应变传感器在简单流动模型中进行无创流量监测

阅读:9
作者:Sunghun Jung, Dae Yu Kim

Abstract

In this paper, we introduce a monitoring method for flow expansion and contraction in a simple flow phantom based on electrical resistance changes in an epidermal strain sensor attached to the phantom. The flow phantom was fabricated to have a nonflat surface and small modulus that are analogous to human skin. The epidermal sensors made of polydopamine and polyvinyl alcohol show sufficient linearity (R = 0.9969), reproducibility, and self-adhesion properties, as well as high sensitivity to small modulus measurements (<1% tensile strain). Pulsatile flow monitoring experiments were performed by placing the epidermal sensor on the flow phantom and measuring the relative changes in resistance by the heartbeat. Experiments were conducted for three types of vessel diameters (1.5, 2, and 3 mm). In each of the experiments, the vessels were divided into Top, Middle, and Bottom positions. Experiments for each position show that the relative changes in resistance increase proportionally with the diameter of the vessel. The vessels located close to the epidermal layer have greater relative electrical changes. The results were analyzed using the Bernoulli equation and hoop stress formula. This study demonstrates the feasibility of a noninvasive flow monitoring method using a novel resistive strain sensor.