Quantitative, real-time scintillation imaging for experimental comparison of different dose and dose rate estimations in UHDR proton pencil beams

用于实验比较超高剂量率质子笔形束中不同剂量和剂量率估算的定量实时闪烁成像

阅读:1

Abstract

BACKGROUND: Ultra-high dose rate radiotherapy (UHDR-RT) has demonstrated normal tissue sparing capabilities, termed the FLASH effect; however, available dosimetry tools make it challenging to characterize the UHDR beams with sufficiently high concurrent spatial and temporal resolution. Novel dosimeters are needed for safe clinical implementation and improved understanding of the effect of UHDR-RT. PURPOSE: Ultra-fast scintillation imaging has been shown to provide a unique tool for spatio-temporal dosimetry of conventional cyclotron pencil beam scanning (PBS) deliveries, indicating the potential use for characterization of UHDR PBS proton beams. The goal of this work is to introduce this novel concept and demonstrate its capabilities in recording high-resolution dose rate maps at FLASH-capable proton beam currents, as compared to log-based dose rate calculation, internally developed UHDR beam simulation, and a fast point detector (EDGE diode). METHODS: The light response of a scintillator sheet located at isocenter and irradiated by PBS proton fields (40-210 nA, 250 MeV) was imaged by an ultra-fast iCMOS camera at 4.5-12 kHz sampling frequency. Camera sensor and image intensifier gain were optimized to maximize the dynamic range; the camera acquisition rate was also varied to evaluate the optimal sampling frequency. Large field delivery enabled flat field acquisition for evaluation of system response homogeneity. Image intensity was calibrated to dose with film and the recorded spatio-temporal data was compared to a PPC05 ion chamber, log-based reconstruction, and EDGE diode. Dose and dose rate linearity studies were performed to evaluate agreement under various beam conditions. Calculation of full-field mean and PBS dose rate maps were calculated to highlight the importance of high resolution, full-field information in UHDR studies. RESULTS: Camera response was linear with dose (R(2) = 0.997) and current (R2(2) = 0.98) in the range from 2-22 Gy and 40-210 nA, respectively, when compared to ion chamber readings. The deviation of total irradiation time calculated with the imaging system from the log file recordings decreased from 0.07% to 0.03% when imaging at 12 kfps versus 4.5 kfps. Planned and delivered spot positions agreed within 0.2 ± 0.1 mm and total irradiation time agreed within 0.2 ± 0.2 ms when compared with the log files, indicating the high concurrent spatial and temporal resolution. For all deliveries, the PBS dose rate measured at the diode location agreed between the imaging and the diode within 3% ± 2% and with the simulation within 5% ± 3% CONCLUSIONS: Full-field mapping of dose and dose rate is imperative for complete understanding of UHDR PBS proton dose delivery. The high linearity and various spatiotemporal metric reporting capabilities confirm the continued use of this camera system for UHDR beam characterization, especially for spatially resolved dose rate information.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。