Impact of low-amplitude pulse interspersed with histotripsy treatment of Staphylococcus aureus biofilms on surgical mesh

低振幅脉冲与组织碎裂术联合治疗金黄色葡萄球菌生物膜对外科网片的影响

阅读:1

Abstract

Ultrasound histotripsy has shown great potential for destroying bacteria biofilms on hernia mesh. The therapy could be further enhanced by reducing the necessary treatment times. In this study, we evaluated the inclusion of a low-amplitude pulse spaced between each of the high-amplitude histotripsy pulses on the effectiveness of the therapy for different treatment durations. The low-amplitude pulse consisted of between 20 and 500 cycles at 1.05 MHz with rarefactional pressure amplitudes varying from 0 (no low-amplitude pulse) to -1.9 MPa. The scan speed was varied as 0.2 mm/s, 0.8 mm/s, and 1.5 mm/s (0.18 BW/s, 0.72 BW/s, 1.35 BW/s where BW is beamwidth). The step size between scan lines was 0.2 mm or 0.4 mm (0.18 BW and 0.36 BW). The pulse repetition frequency (PRF) was varied between 62.5 and 500 Hz with the low-amplitude pulses (if present) positioned 500μs after the histotripsy pulses. The number of colony-forming units on the mesh samples was significantly reduced when a low-amplitude pulse was included between the therapy pulses provided the mechanical index (MI) of the low-amplitude pulse exceeded 1.5. Slower scan speeds also increased biofilm destruction, but the effect was much less pronounced except for very slow scan speeds of 0.18 BW/s. Thus, the use of secondary pulses reduced treatment times from 20.9 min/cm(2) to 2.78 min/cm(2). Varying the PRF had no impact when the secondary pulse was not included, but biofilm destruction increased at higher PRFs when a secondary pulse was present. The number of cycles in the secondary pulse also had little impact over the range of values tested.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。