Performance evaluation of the sterilization process with Bowie & Dick test and biological indicator in the quality control of a blood bank in Peru

在秘鲁某血库的质量控制中,采用鲍伊-迪克试验和生物指示剂对灭菌过程进行性能评估。

阅读:1

Abstract

Managing contaminated waste in blood banks is a crucial process that must end with the safe disposal of blood products. In this sense, practical methods and indicators must be used to guarantee the operability of autoclave equipment and the sterility of treated waste to mitigate the risk of biological exposure and hospital biosafety. We designed a cross-sectional study to evaluate the performance the autoclaving process with Bowie & Dick test and biological indicator at the Hemotherapy and Blood Bank Unit of the Cayetano Heredia Hospital in Peru. Fifty autoclaving processes were carried out independently for the Bowie & Dick tests and biological indicators based on Geobacillus stearothermophilus spores. Autoclaving was programmed at 134°C for 3.5 minutes for the Bowie & Dick test, while, for the biological indicator, it was programmed at 121°C for 5 minutes. Both in the presence of contaminated waste. The autoclaving process evaluated by the Bowie & Dick test had a compliance rate of 80% (CI95: 66.3%-90.0%), while, by the biological indicator, 90% (CI95: 78.2%-96.7%). We did not find significant differences between the results of both tests (P = .689; Fisher exact test). The compliance rate in the autoclaving process within the blood bank of a Peruvian Hospital is acceptable; however, there are actions for continuous improvement, especially at the vacuum level in the autoclaving process.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。