Comparative analysis of the mechanical limits of resistance in implant/abutment set of a new implant design: An in vitro study

新型种植体设计种植体/基台组件机械强度极限的比较分析:一项体外研究

阅读:1

Abstract

OBJECTIVE: The aim of the present in vitro study was to evaluate the resistance on quasi-static forces and in the fatigue mechanical cycling of a new implant design compared to two other conventional implant designs. MATERIALS AND METHODS: Eighty-eight implants with their respective abutments were tested and distributed into four groups (n = 22 per group): Morse taper connection implant (MT group), conventional external hexagon implant (EH con group), new Collo implant of external hexagon with the smooth portion out of the bone insertion (EH out group), and new Collo implant of external hexagon with the implant platform inserted to the bone level (EH bl group). All the sets were subjected to quasi-static loading in a universal testing machine, and we measured the maximum resistance force supported by each sample. Another 12 samples from each group were submitted to the cyclic fatigue test at 4 intensities of forces (n = 3 per force): 80%, 60%, 40%, and 20%. The number of cycles supported by each sample at each force intensity was evaluated. RESULTS: The three groups of implants with external hexagon connection had similar maximum strength values of the sets (p > 0.05). Meanwhile, samples from the MT group showed the highest resistance values in comparison to the other three groups (p < 0.05). In the fatigue test, the Collo out group supported a smaller number of cycles that led to the fracture than the other 3 groups proposed at loads of 80%, 60%, and 40%, and only at the load value of 20% all groups had the same performance. CONCLUSIONS: Within the limitations of the present in vitro study, the results showed that the new Collo implant performs better when installed at bone level.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。