Optimization of Image Quality and Organ Absorbed Dose for Pediatric Chest X-Ray Examination: In-House Developed Chest Phantom Study

儿童胸部X光检查图像质量和器官吸收剂量的优化:自主研发的胸部模型研究

阅读:1

Abstract

PURPOSE: This study aimed to identify proper exposure techniques to maintain optimal diagnostic image quality with minimum radiation dose for anteroposterior chest X-ray projection in pediatric patients. METHODS: Briefly, an in-house developed pediatric chest phantom was constructed. Next, nanodot OSLDs were used for organ absorbed dose measurement and placed in the lung area, and the phantom was exposed to various exposure techniques (ranging from 50 to 70 kVp with 1.6, 2, and 2.5 mAs). After that, the phantom was used to assess image quality parameters, including SNR and CNR. Two radiologists assessed the subjective image quality using a visual grading analysis (VGA) technique. Finally, the figure of merit (FOM) was analyzed. RESULTS: The developed phantom was constructed successfully and could be useful for dose measurement and image quality assessment. The absorbed dose varied from 0.009 to 0.031 mGy for the range of exposure techniques used. SNR and CNR showed a gradually increasing trend, while kVp and mAs values were increased. The highest kVp (70 kVp) produced the highest SNR and CNR, exhibiting a significant difference compared with 50 and 60 kVp (P < 0.05). The overall VGA score was 3.2 ± 0.3, and the low kVp technique demonstrated better image quality compared with the reference image. CONCLUSION: The optimized exposure technique was identified as 60 kV and 2.5 mAs, indicating the highest FOM score. This work revealed practicable techniques that could be implemented into clinical practice for performing pediatric chest radiography.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。