Effective Cardiopulmonary Resuscitation (CPR) requires precise chest compression depth, but current out-of-hospital monitoring technologies face limitations. This study introduces a method using frequency-modulated continuous-wave (FMCW) radar to remotely and accurately monitor chest compressions. FMCW radar captures range, Doppler, and angular data, and we utilize micro-Doppler signatures for detailed motion analysis. By integrating Doppler shifts over time, chest displacement is estimated. We compare a regression model based on maximum Doppler frequency with deep convolutional neural networks (DCNNs) trained on spectrograms generated via short-time Fourier transform (STFT) and the Wigner-Ville distribution (WVD). The regression model achieved a root mean square error (RMSE) of 0.535 cm. The STFT-based DCNN improved accuracy with an RMSE of 0.505 cm, while the WVD-based DCNN achieved the best performance with an RMSE of 0.447 cm, representing an 11.5% improvement over the STFT-based DCNN. These findings highlight the potential of combining FMCW radar and deep learning to provide accurate, real-time chest compression depth measurement during CPR, supporting the development of advanced, non-contact monitoring systems for emergency medical response.