Measurement Device for Ambient Carbonyl Sulfide by Means of Catalytic Reduction Followed by Wet Scrubbing/Fluorescence Detection

一种利用催化还原结合湿式洗涤/荧光检测法测量环境羰基硫的装置

阅读:1

Abstract

A portable chemical analysis system for monitoring ambient carbonyl sulfide (COS) was investigated for the first time. COS is paid attention to from the perspectives of photosynthesis tracer, breath diagnosis marker, and new process-use in the manufacture of semiconductors. Recently, the threshold level value of COS was settled at 5 ppm in volume ratio (ppmv) for workplace safety management. In this work, COS was converted to H(2)S by a small column packed with alumina catalyzer at 65 °C. Then, the H(2)S produced was collected in a small channel scrubber to react with fluorescein mercuric acetate (FMA), and the resulting fluorescence quenching was monitored using an LED/photodiode-based miniature detector. The miniature channel scrubber was re-examined to determine its robustness and easy fabrication, and conditions of the catalyzer were optimized. When the FMA concentration used was 1 μM, the limit of detection and dynamic range, which were both proportional to the FMA concentration, were 0.07 and 25 ppbv, respectively. Ambient COS in the background level and even contaminated COS in the nitrogen gas cylinder could be detected. If necessary, H(2)S was removed selectively by reproducible adsorbent columns. COS concentrations of engine exhaust were measured by the proposed method and by cryo-trap-gas chromatography-flame photometric detection, and the results obtained (0.5-5.9 ppbv) by the two methods agreed well (R (2) = 0.945, n = 19). COS in ambient air and exhaust gases was successfully measured without any batchwise pretreatment.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。