Comparative validation of a microcapsule-based immunoassay for the detection of proteins and nucleic acids

基于微胶囊的免疫测定法检测蛋白质和核酸的比较验证

阅读:7
作者:Sujit Kumar Verma, Anja Karin Albrecht, Verena Siebecke, Gerd Klöck, Tatiana A Kolesnikova, Sebastian Springer

Abstract

To detect and study diseases, research and clinical laboratories must quantify specific biomarkers in the plasma and urine of patients with precision, sensitivity, and cost-effectiveness. Newly developed techniques, such as particle-based immunoassays, must be validated in these terms against standard methods such as enzyme-linked immunosorbent assays (ELISAs). Here, we compare the performance of assays that use hollow polyelectrolyte microcapsules with assays based on solid plastic beads, and with standard microplate immunoassays. The polyelectrolyte microcapsules detect the disease biomarker beta-2 microglobulin with a fifty-fold increase in sensitivity than polystyrene (PS) beads. For sequence-specific nucleic acid detection, the oligonucleotide-coated microcapsules exhibit a two-fold lower increase in sensitivity over PS beads. The microcapsules also detect the presence of a monoclonal antibody in hybridoma supernatant at a fifty-six-fold increase in sensitivity compared to a microplate assay. Overall, polyelectrolyte microcapsule-based assays are more sensitive for the detection of protein and nucleic acid analytes than PS beads and microplate assays, and they are viable alternatives as a platform for the rapid quantitative detection of analytes at very low concentrations.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。