Optimization of Regularization Parameters in Compressed Sensing of Magnetic Resonance Angiography: Can Statistical Image Metrics Mimic Radiologists' Perception?

磁共振血管造影压缩感知中正则化参数的优化:统计图像指标能否模拟放射科医生的感知?

阅读:1

Abstract

In Compressed Sensing (CS) of MRI, optimization of the regularization parameters is not a trivial task. We aimed to establish a method that could determine the optimal weights for regularization parameters in CS of time-of-flight MR angiography (TOF-MRA) by comparing various image metrics with radiologists' visual evaluation. TOF-MRA of a healthy volunteer was scanned using a 3T-MR system. Images were reconstructed by CS from retrospectively under-sampled data by varying the weights for the L1 norm of wavelet coefficients and that of total variation. The reconstructed images were evaluated both quantitatively by statistical image metrics including structural similarity (SSIM), scale invariant feature transform (SIFT) and contrast-to-noise ratio (CNR), and qualitatively by radiologists' scoring. The results of quantitative metrics and qualitative scorings were compared. SSIM and SIFT in conjunction with brain masks and CNR of artery-to-parenchyma correlated very well with radiologists' visual evaluation. By carefully selecting a region to measure, we have shown that statistical image metrics can reflect radiologists' visual evaluation, thus enabling an appropriate optimization of regularization parameters for CS.