Effect of optical blurring of X-ray source on breast tomosynthesis image quality: Modulation transfer function, anatomical noise power spectrum, and signal detectability perspectives

X射线源光学模糊对乳腺断层合成图像质量的影响:调制传递函数、解剖噪声功率谱和信号可检测性视角

阅读:1

Abstract

We investigated the effect of the optical blurring of X-ray source on digital breast tomosynthesis (DBT) image quality using well-designed DBT simulator and table-top experimental systems. To measure the in-plane modulation transfer function (MTF), we used simulated sphere phantom and Teflon sphere phantom and generated their projection data using two acquisition modes (i.e., step-and-shoot mode and continuous mode). After reconstruction, we measured the in-plane MTF using reconstructed sphere phantom images. In addition, we measured the anatomical noise power spectrum (aNPS) and signal detectability. We constructed simulated breast phantoms with a 50% volume glandular fraction (VGF) of breast anatomy using the power law spectrum and inserted spherical objects with 1 mm, 2 mm, and 5 mm diameters as breast masses. Projection data were acquired using two acquisition modes, and in-plane breast images were reconstructed using the Feldkamp-Davis-Kress (FDK) algorithm. For the experimental study, we used BR3D breast phantom with 50% VGF and obtained projection data using a table-top experimental system. To compare the detection performance of the two acquisition modes, we calculated the signal detectability using the channelized Hotelling observer (CHO) with Laguerre-Gauss (LG) channels. Our results show that spatial resolution of in-plane image in continuous mode was degraded due to the optical blurring of X-ray source. This blurring effect was reflected in aNPS, resulting in large β values. From a signal detectability perspective, the signal detectability in step-and-shoot mode is higher than that in continuous mode for small spherical signals but not large spherical signals. Although the step-and-shoot mode has disadvantage in terms of scan time compared to the continuous mode, scanning in step-and-shoot mode is better for detecting small signals, indicating that there is a tradeoff between scan time and image quality.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。