Application of Excimer Lamp in Quantitative Detection of SF(6) Decomposition Component SO(2)

准分子灯在SF(6)分解产物SO(2)定量检测中的应用

阅读:1

Abstract

Accurate quantitative detection for trace gas has long been the center of failure diagnosis for gas-insulated equipment. An absorption spectroscopy-based detection system was developed for trace SF(6) decomposition SO(2) detection in this paper. In order to reduce interference from other decomposition, ultraviolet spectrum of SO(2) was selected for detection. Firstly, an excimer lamp was developed in this paper as the excitation of the absorption spectroscopy compared with regular light sources with electrodes, such as electrodeless lamps that are more suitable for long-term monitoring. Then, based on the developed excimer lamp, a detection system for trace SO(2) was established. Next, a proper absorption peak was selected by calculating spectral derivative for further analysis. Experimental results indicated that good linearity existed between the absorbance and concentration of SO(2) at the chosen absorption peak. Moreover, the detection limit of the proposed detection system could reach the level of 10(-7). The results of this paper could serve as a guide for the application of excimer lamp in online monitoring for SF(6)-insulated equipment.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。