A Simple AMC Antenna for Liquid Monitoring in an Infusion Bag

一种用于输液袋液体监测的简易型AMC天线

阅读:1

Abstract

Running-out detection of the liquids in an infusion bag is important for medical treatment. This paper proposed a simple low-cost sensing scheme with an artificial magnetic conductor (AMC) antenna for liquid-running-out detection in infusion bags. The proposed antenna consists of a dipole antenna supported by an AMC layer. It operates in the 2.4 GHz ISM band in the without-liquid state, in the 2.0 GHz ISM band in the with-liquid state, and can be used for liquid sensing. The AMC layer isolates interference from the surrounding environment such as the standing pole. It also enhances antenna performance and improves monitoring sensitivity. This gives a peak gain of 6.45 dBi and a radiation efficiency of 98% in the without-liquid state. Meanwhile, the with-liquid state can achieve a peak gain of 4.5 dBi and a radiation efficiency of 93%. The proposed antenna is fabricated and measured, verifying its sensing performance of the liquid in the infusion bag. This antenna's design is flexible, compact, precise, and suitable for biomedical wireless sensing.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。