Performance Characterization of a Fully Transportable Mid-Infrared Laser Heterodyne Radiometer (LHR)

全便携式中红外激光外差辐射计(LHR)的性能表征

阅读:1

Abstract

A fully transportable laser heterodyne radiometer (LHR), involving a flexible polycrystalline mid-infrared (PIR) fiber-coupling system and operating around 8 µm, was characterized and optimized with the help of a calibrated high temperature blackbody source to simulate solar radiation. Compared to a mid-IR free-space sunlight coupling system, usually used in a current LHR, such a fiber-coupling system configuration makes the mid-infrared (MIR) LHR fully transportable. The noise sources, heterodyne signal, and SNR of the MIR LHR were analyzed, and the optimum operating local oscillator (LO) photocurrent was experimentally obtained. The spectroscopic performance of the MIR LHR was finally evaluated. This work demonstrated that the developed fully transportable MIR LHR could be used for ground-based atmospheric sounding measurements of multiple trace gases in the atmospheric column. In addition, it also has high potential for applications on spacecraft or on an airborne platform.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。