Novel Microfluidic Analytical Sensing Platform for the Simultaneous Detection of Three Algal Toxins in Water

一种用于同时检测水中三种藻毒素的新型微流控分析传感平台

阅读:1

Abstract

Globally, the need for "on-site" algal-toxin monitoring has become increasingly urgent due to the amplified demand for fresh-water and for safe, "toxin-free" shellfish and fish stocks. Herein, we describe the first reported, Lab-On-A-Disc (LOAD) based-platform developed to detect microcystin levels in situ, with initial detectability of saxitoxin and domoic acid also reported. Using recombinant antibody technology, the LOAD platform combines immunofluorescence with centrifugally driven microfluidic liquid handling to achieve a next-generation disposable device capable of multianalyte sampling. A low-complexity "LED-photodiode" based optical sensing system was tailor-made for the platform, which allows the fluorescence signal of the toxin-specific reaction to be quantified. This system can rapidly and accurately detect the presence of microcystin-LR, domoic acid, and saxitoxin in 30 min, with a minimum of less than 5 min end-user interaction for maximum reproducibility. This method provides a robust "point of need" diagnostic alternative to the current laborious and costly methods used for qualitative toxin monitoring.