Feasibility Study of a Time-of-Flight Brain Positron Emission Tomography Employing Individual Channel Readout Electronics

采用独立通道读出电子设备的飞行时间脑正电子发射断层扫描可行性研究

阅读:1

Abstract

The purpose of this study was to investigate the feasibility of a time-of-flight (TOF) brain positron emission tomography (PET) providing high-quality images. It consisted of 30 detector blocks arranged in a ring with a diameter of 257 mm and an axial field of view of 52.2 mm. Each detector block was composed of two detector modules and two application-specific integrated circuit (ASIC) chips. The detector module was composed of an 8 × 8 array of 3 × 3 mm(2) multi-pixel photon counters and an 8 × 8 array of 3.11 × 3.11 × 15 mm(3) lutetium yttrium oxyorthosilicate scintillators. The 64-channel individual readout ASIC was used to acquire the position, energy, and time information of a detected gamma ray. A coincidence timing resolution of 187 ps full width at half maximum (FWHM) was achieved using a pair of channels of two detector modules. The energy resolution and spatial resolution were 6.6 ± 0.6% FWHM (without energy nonlinearity correction) and 2.5 mm FWHM, respectively. The results of this study demonstrate that the developed TOF brain PET could provide excellent performance, allowing for a reduction in radiation dose or scanning time for brain imaging due to improved sensitivity and signal-to-noise ratio.