Application of Mixed Effects Limits of Agreement in the Presence of Multiple Sources of Variability: Exemplar from the Comparison of Several Devices to Measure Respiratory Rate in COPD Patients

在存在多种变异来源的情况下应用混合效应一致性界限:以几种测量慢性阻塞性肺病患者呼吸频率的设备比较为例

阅读:1

Abstract

INTRODUCTION: The Bland-Altman limits of agreement method is widely used to assess how well the measurements produced by two raters, devices or systems agree with each other. However, mixed effects versions of the method which take into account multiple sources of variability are less well described in the literature. We address the practical challenges of applying mixed effects limits of agreement to the comparison of several devices to measure respiratory rate in patients with chronic obstructive pulmonary disease (COPD). METHODS: Respiratory rate was measured in 21 people with a range of severity of COPD. Participants were asked to perform eleven different activities representative of daily life during a laboratory-based standardised protocol of 57 minutes. A mixed effects limits of agreement method was used to assess the agreement of five commercially available monitors (Camera, Photoplethysmography (PPG), Impedance, Accelerometer, and Chest-band) with the current gold standard device for measuring respiratory rate. RESULTS: Results produced using mixed effects limits of agreement were compared to results from a fixed effects method based on analysis of variance (ANOVA) and were found to be similar. The Accelerometer and Chest-band devices produced the narrowest limits of agreement (-8.63 to 4.27 and -9.99 to 6.80 respectively) with mean bias -2.18 and -1.60 breaths per minute. These devices also had the lowest within-participant and overall standard deviations (3.23 and 3.29 for Accelerometer and 4.17 and 4.28 for Chest-band respectively). CONCLUSIONS: The mixed effects limits of agreement analysis enabled us to answer the question of which devices showed the strongest agreement with the gold standard device with respect to measuring respiratory rates. In particular, the estimated within-participant and overall standard deviations of the differences, which are easily obtainable from the mixed effects model results, gave a clear indication that the Accelerometer and Chest-band devices performed best.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。