Design and quantitative evaluation of 'Aerosol Bio-Containment Device (ABCD)' for reducing aerosol exposure during infectious aerosol-generating events

设计和定量评估“气溶胶生物隔离装置(ABCD)”,以减少传染性气溶胶生成事件期间的气溶胶暴露。

阅读:4

Abstract

The Coronavirus Disease 2019 (COVID-19) pandemic renewed interest in infectious aerosols and reducing risk of airborne respiratory pathogen transmission, prompting development of devices to protect healthcare workers during airway procedures. However, there are no standard methods for assessing the efficacy of particle containment with these protective devices. We designed and built an aerosol bio-containment device (ABCD) to contain and remove aerosol via an external suction system and tested the aerosol containment of the device in an environmental chamber using a novel, quantitative assessment method. The ABCD exhibited a strong ability to control aerosol exposure in experimental and computational fluid dynamic (CFD) simulated scenarios with appropriate suction use and maintenance of device seals. Using a log-risk-reduction framework, we assessed device containment efficacy and showed that, when combined with other protective equipment, the ABCD can significantly reduce airborne clinical exposure. We propose this type of quantitative analysis serves as a basis for rating efficacy of aerosol protective enclosures.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。