Development and Validation of a Portable EIT System for Real-Time Respiratory Monitoring

便携式EIT系统在实时呼吸监测中的开发与验证

阅读:1

Abstract

This work contributes to the improvement of novel medical technologies for the prevention and treatment of diseases. Electrical impedance tomography (EIT) has gained attention as a valuable tool for non-invasive monitoring providing real-time insights. The purpose of this work is to develop and validate a novel portable EIT system with a small form factor for respiratory monitoring. The device uses a 16-electrode architecture with adjacent stimulation and measurement patterns, an integrated circuit current source and a single high-speed ADC operating with multiplexers to stimulate and measure across all electrodes. Tests were conducted on 25 healthy subjects who performed a pulmonary function test with a flowmeter while using the EIT device. The results showed a good performance of the device, which was able to recognize all respirations correctly, and from the EIT signals and images, correlations of 96.7% were obtained for instantaneous respiratory rate and 96.1% for tidal volume prediction. These results validate the preliminary technical feasibility of the EIT system and demonstrates its potential as a reliable tool for non-invasive respiratory assessment. The significance of this work lies in its potential to democratize advanced respiratory monitoring technologies, making them accessible to a wider population, including those in remote or underserved areas.