Background

Haemophilus influenzae was the most aggressive pathogen and formed a major cause of bacterial meningitis and pneumonia in young children and infants, which need medical emergency requiring immediate diagnosis and treatment. However, From isolation to identification of H. influenzae, the traditional diagnose strategy was time-consuming and expensive. Therefore, the establishment of a convenient, highly sensitive, and stable detection system is urgent and critical.

Conclusions

Taken together, LAMP-LFB could be used as an effective diagnostic approach for H. influenzae in the conditions of basic and clinical labs, which would allow clinicians to make better informed decisions regarding patient treatment without delay.

Results

In this study, we used a combined method to detect H. influenzae. Six specific primers were designed on the basis of outer membrane protein P6 gene sequence of H. influenzae. The reaction condition such as the optimum temperature was 65℃, and the optimum reaction time was 30 min, respectively. Through the loop-mediated isothermal amplification (LAMP) in combination with nanoparticle-based lateral flow biosensor (LFB), the sensitivity of LAMP-LFB showed 100 fg was the lowest genomic DNA templates concentration in the pure cultures. Meanwhile, the specificity of H. influenzae-LAMP-LFB assay showed the exclusive positive results, which were detected in H. influenzae templates. In 55 clinical sputum samples, 22 samples were positive with LAMP-LFB method, which was in accordance with the traditional culture and Polymerase Chain Reaction (PCR) method. The accuracy in diagnosing H. influenzae with LAMP-LFB could reach 100%, compared to culture and PCR method, indicating the LAMP-LFB had more advantages in target pathogen detection. Conclusions: Taken together, LAMP-LFB could be used as an effective diagnostic approach for H. influenzae in the conditions of basic and clinical labs, which would allow clinicians to make better informed decisions regarding patient treatment without delay.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。