Factors influencing the reliability of measurements in eyes with full-thickness macular holes: are we measuring incorrectly?

影响全层黄斑裂孔眼测量可靠性的因素:我们的测量方法是否不正确?

阅读:2

Abstract

PURPOSE: The calliper function is used for manual measurements of full thickness macular holes (FTMHs). We aimed to investigate whether a reproducible difference could be detected beyond interobserver variability between two commonly used manufacturers in their manual calliper facility in spectral domain optical coherence tomography (OCT) for metrics related to FTMH. METHODS: This is a non-interventional, retrospective, observational study. Two independent observers examined 8 eyes (16 OCT) scans and 128 measurements (minimal linear diameter (MLD), basal diameter and hole height on both sides) of FTMHs, taken on Heidelberg Spectralis and Topcon Triton (OCT machines). The interobserver agreement and OCT machine agreement of measurements were analysed by Bland-Altman plots and intraclass correlation coefficient (ICC) analysis. Spectralis and Triton had 125 µm and 50 µm horizontal b-scan spacing, respectively. RESULTS: Overall, we report high absolute agreement in interobserver (ICC 0.991 (95% CI 0.985 to 0.995, p<0.001)) and OCT machine (ICC 0.993 (95% CI 0.987 to 0.996, p<0.001)) variability. Lower horizontal resolution in Triton compared with Spectralis leads to interobserver variability, in smaller horizontal measurements. Lower horizontal scanning density in Spectralis lead to relatively large interobserver variation if different reference scans were chosen, and consistently smaller MLD measurements than Triton. Vertical measurements without 1:1 scaling lead to inaccurate exaggerated oblique vertical measurements. Calliper function appears otherwise identically calibrated. CONCLUSIONS: We report excellent interobserver and OCT machine agreement in measurements. However, the paper shows several factors that could influence the reliability of measurements acquired in eyes with FTMHs, such as the dimension of the hole as well as different image resolution, density scanning protocols or vertical scaling of the OCT machines viewing platform.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。