The Impact of Air Inflow and Interfering Factors on the Performance of Microbiological Safety Cabinets

空气流入和干扰因素对微生物安全柜性能的影响

阅读:1

Abstract

INTRODUCTION: The operator protection factor (OPF) of four biological safety cabinets (BSCs) has been measured under standard and suboptimal conditions. METHODS: The OPF for one BSC1, two BSC2, and an acid-fast bacilli staining station (AFBSS) was measured using the potassium iodide method for in situ testing of BSCs (CEN12469) over a range of inflow velocities under standard conditions and with common interfering factors (fans, opening doors, and walk pasts). RESULTS: The BSC1 and the AFBSS gave a high level of protection under standard test conditions at all airflows (down to 0.3 and 0.38 m/s, respectively). During interfering processes, the BSC1 and AFBSS gave a high level of protection (OPF >10(5)) at the specified inward airflow. At lower airflows, there was a predictable deterioration in performance. There was a significant difference in performance between the two BSC2s tested, with one model passing all tests under all interfering conditions at all airflows. The second BSC2 failed the standard test at the lowest airflow and provided poor levels of protection (OPF <10(5)) in all tests carried out with interfering processes. CONCLUSION: Although BSC2s are capable of giving a high level of performance, this is design dependent and the BSC1 and AFBSS give a more predictable level of performance due to their simpler design. In environments where BSC certification is not possible, they may provide more robust and sustainable primary containment.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。