Endoscopy-MR Image Fusion for Image Guided Procedures

内镜-磁共振图像融合在图像引导手术中的应用

阅读:1

Abstract

Minimally invasive endoscope based abdominal procedures provide potential advantages over conventional open surgery such as reduced trauma, shorter hospital stay, and quick recovery. One major limitation of using this technique is the narrow view of the endoscope and the lack of proper 3D context of the surgical site. In this paper, we propose a rapid and accurate method to align intraoperative stereo endoscopic images of the surgical site with preoperative Magnetic Resonance (MR) images. Gridline light pattern is projected on the surgical site to facilitate the registration. The purpose of this surface-based registration is to provide 3D context of the surgical site to the endoscopic view. We have validated the proposed method on a liver phantom and achieved the surface registration error of 0.76 ± 0.11 mm.