Enhancement of human plasma glucosylceramide assay sensitivity using delipidized plasma

利用脱脂血浆提高人血浆葡糖基神经酰胺测定的灵敏度

阅读:1

Abstract

Glucosylceramide (GL-1) level in human has been considered as a surrogate biomarker for enzyme replacement and substrate reduction therapies (ERT and SRT) for Gaucher and Fabry patients. Due to the high endogenous level of GL-1 in human plasma, it is difficult to achieve the analytical sensitivity of plasma GL-1 below the normal endogenous level (1.7 μg/mL to 6.6 μg/mL) when using the standard addition method and regular plasma matrix for standard curve. A high sensitivity plasma GL-1 assay with LLOQ at 0.1 μg/mL was developed and validated using delipidized plasma so that patient plasma concentrations that are below normal reference range can be measured accurately. The normal reference range was established from 120 healthy donors using this developed new method. Twenty-three Fabry patient plasma samples including baseline and post-investigation drug treatment samples were measured. All post-treatment samples showed GL-1 concentration below 2.0 μg/mL, indicating the utility of the reported high sensitivity assay using delipidized plasma for monitoring the plasma GL-1 biomarker level in patients.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。