A Novel Reduced Graphene Oxide Modified Carbon Paste Electrode for Potentiometric Determination of Trihexyphenidyl Hydrochloride in Pharmaceutical and Biological Matrices

新型还原氧化石墨烯修饰碳糊电极用于药物和生物基质中盐酸苯海索的电位测定

阅读:10
作者:Josip Radić, Maša Buljac, Boštjan Genorio, Ema Gričar, Mitja Kolar

Abstract

A novel promising carbon paste electrode with excellent potentiometric properties was prepared for the analysis of trihexyphenidyl hydrochloride (THP), the acetylcholine receptor and an anticholinergic drug in real samples. It contains 10.2% trihexyphenidy-tetraphenylborate ionic pair as the electroactive material, with the addition of 3.9% reduced graphene oxide and 0.3% of anionic additive into the paste, which consists of 45.0% dibutylphthalate as the solvent mediator and 40.6% graphite. Under the optimized experimental conditions, the electrode showed a Nernstian slope of 58.9 ± 0.2 mV/decade with a regression coefficient of 0.9992. It exhibited high selectivity and reproducibility as well as a fast and linear dynamic response range from 4.0 × 10-7 to 1.0 × 10-2 M. The electrode remained usable for up to 19 days. Analytical applications showed excellent recoveries ranging from 96.8 to 101.7%, LOD was 2.5 × 10-7 M. The electrode was successfully used for THP analysis of pharmaceutical and biological samples.